ISO class from particle counts: the guide

A cleanroom's class is a particle count and nothing else; the calculator turns the counter's reading into the ISO class and shows how much room is left under the limit.

Open the calculator

Why this number matters

ISO 14644-1 defines nine classes by the number of particles of a given size allowed per cubic meter of air, on a formula that scales smoothly with size. At 0.5 micrometers, the size most rooms are classified on, ISO 5 allows 3,520, ISO 6 35,200, ISO 7 352,000, and ISO 8 3,520,000. The old Federal Standard 209E names, Class 100, Class 10,000, Class 100,000, are the same limits counted per cubic foot, and they are still what most people say.

Classification is done with a laser particle counter at a number of sample points that depends on the floor area, at work height, with the room in the state you claim, at rest or in operation. The room meets the class if every size you measured is under that class's limit at every point, with the statistical allowance the standard sets. Counts are also the leading indicator of trouble: a room that runs at 60% of its limit on a good day fails on the day the HEPA filter loads or a gowning rule slips, which is why the headroom number matters as much as the class.

The class drives the design: air changes, ceiling coverage with fan filter units, pressure cascade, gowning, and the materials and finishes. ISO 8 is a packaging and assembly room; ISO 7 is medical device and electronics work; ISO 5 is unidirectional flow over the critical process, usually inside an ISO 7 room.

How it is figured

  • ISO 14644-1 class limits: Cn = 10^N × (0.1 ÷ D)^2.08 particles per cubic meter at size D in µm, so at 0.5 µm ISO 5 is 3,520, ISO 6 35,200, ISO 7 352,000, ISO 8 3,520,000. The class is the smallest N whose limit your count stays under at every size you measured.
  • Counters that read per cubic foot are converted at 35.3147 ft³ per m³. Federal Standard 209E classes (Class 100, 10,000, 100,000) are the particles per ft³ at 0.5 µm and map to ISO 5, 7, and 8.

A worked example

A counter reading 120,000 particles per cubic meter at 0.5 micrometers is under the ISO 7 limit of 352,000 with 66% headroom; the room is ISO 7, a Class 10,000 in the old terms, and would need two-thirds fewer particles to reach ISO 6.

The tool's starting values, and what it returns for them:

Counter reading percubic meter (ISO)
Particles at 0.5 µm and larger120,000
At 0.1 µm (0 if not measured)0
At 0.3 µm (0 if not measured)0
At 1 µm (0 if not measured)0
At 5 µm (0 if not measured)0
Room meetsISO Class 7
ISO 7 limit at 0.5 µm352,000 per m³
Headroom under the limit66 %
ISO 6 limit at 0.5 µm35,200 per m³
ISO 5 limit at 0.5 µm3,520 per m³
ISO 635,200 per m³
ISO 7352,000 per m³
ISO 83,520,000 per m³
Exact class number6.53

How to count

  1. Run the room at rest or in operation (say which), HEPA units on, doors closed, for at least 15 minutes.
  2. Sample with a calibrated particle counter at the number of locations ISO 14644-1 Annex A requires for the floor area (√area in m² rounded up, minimum per the table), 1 minute or the volume the class needs, at work height.
  3. Enter the highest location's reading, or the 95% UCL the counter software reports.

Tools: a laser particle counter (0.3 or 0.5 µm channel), calibrated within a year

Mistakes that catch people

  • Classifying on one sample point in the middle of the room; the standard sets the number of points by area.
  • Reading a counter that reports per cubic foot as if it were per cubic meter, which is a factor of 35.
  • Claiming a class at rest and running the process in it as if the class held in operation.

Rules of thumb

  • ISO 5 = Class 100, ISO 7 = Class 10,000, ISO 8 = Class 100,000.
  • Each ISO class is ten times the particles of the class below.
  • Run at 50 to 60% of the limit so an ordinary bad day does not fail the room.

When to stop calculating and make a call

  • Certifying a room for a customer or an audit: a certified testing firm does the classification to the standard.
  • A room that passed at rest fails in operation: the people, the process, and the gowning are the investigation, not the filters.

Questions people ask

What ISO class is a Class 10,000 cleanroom?
ISO 7. Class 10,000 meant 10,000 particles of 0.5 µm per cubic foot, which is 352,000 per cubic meter, the ISO 7 limit. Class 100 is ISO 5 and Class 100,000 is ISO 8, the class most modular cleanrooms for packaging, assembly, and medical device work are built to.
Which particle size do I classify on?
Most rooms on 0.5 µm; electronics and pharma rooms also on 0.3 or 0.1. The class is set by the worst size, and ISO 14644-1 limits which sizes may be used for each class (0.1 µm only up to ISO 6, 5 µm only ISO 7 and above).

Standards and sources

  • ISO 14644-1:2015, Classification of air cleanliness by particle concentration (Table 1 and Annex A)
  • Federal Standard 209E (withdrawn 2001, still used in names)

This guide and the tool are planning aids. The equipment maker, the engineer of record, the accountant, or the code official confirms the final number for your building.